Abstract

Time-reversal with Multiple Signal Classification (TR-MUSIC) is an ultrasound imaging algorithm for detecting small targets embedded in a medium. This technique can produce images with subwavelength resolution when the targets are pointlike, and when the number of targets is fewer than the number of transducer elements used to image the medium. In this experimental study, we evaluate the performance of the TR-MUSIC algorithm when the interrogated medium contains extended targets that cannot be considered as point scatterers. We construct tissue-mimicking phantoms embedded with distributed glass spheres. We show that the quality of the phantom images obtained using the TR-MUSIC algorithm decreases with increasing sphere size. However, significant improvement is achieved when the image plane is divided into sub-regions, where each sub-region is imaged separately. The windowed TR-MUSIC algorithm accurately locates the spheres (extended targets), but the images do not provide quantitative information about the shape and reflectivity of the spheres.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.