Abstract

From January 2019 to January 2020, 106 patients (age, 64.8 ± 14.1 years; male, 63.2%) were included to retrospectively investigate the feasibility and safety of ultrasound-guided deployment of ExoSeal after femoral artery access. Baseline characteristics were not different except for age (P = .022), body mass index (P = .009), and diameter (P < .001) between the calcified plaque or stenosis (CS) group (n = 49) and non-CS group (n = 57). The overall technical and clinical success rates were 96.2% and 100%, respectively. The technical (CS group, 48/49; non-CS group, 54/57) and clinical success rates (100%), time to hemostasis (CS group, 3.21 ± 0.54 min; non-CS group, 3.39 ± 0.71 min), and complication rates (CS group, 1/49; non-CS group, 0/57) were not different between the 2 groups. ExoSeal seems to be safe to use under ultrasound guidance in the femoral arteries with CS.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.