In a study of 53 cases, aneurysm size was measured with the use of preoperative physical examination, lateral abdominal roentgenography, and B-mode aortic ultrasonography for comparison with direct intraoperative measurement during aneurysmectomy. Mean aneurysm diameter on the basis of physical examination overestimated aortic size by 1 cm; physical examination measurement was within 0.5 cm of true aortic size in 38% of patients and within 1 cm in 58% of patients. The lateral roentgenogram also overestimated mean aortic diameter by 1 cm and was within 0.5 cm of true aortic size in 44% of patients and within 1 cm in 72% of patients. Measurement using B-mode aortic ultrasonography closely approximated intraoperative aneurysm measurement. Measurement was identical in 34% of patients and was within 0.5 cm and 1 cm in 75% and 92% of patients, respectively.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: JAMA: The Journal of the American Medical Association
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.