Abstract
Problem statement: There is no specific instrument to test ultrasonic dental scalers used in dentistry. Approach: The aim of this research was to develop and test an innovative cavitometer for quantitative and automatic performance assessments of Ultrasonic Dental Scalers (UDS) through the acoustic emission produced by cavitation induced by the tip of the UDS ultrasonic transducer when immersed in aqueous solution. Firstly, an optimized acoustic emission sensor specifically for UDS ultrasonic transducers, software to process the acoustic emission sensor signal and to calculate a Figure of Merit (FM) and a hydraulic-pneumatic circuit were developed; these parts together constituted the cavitometer. To validate the developed cavitometer, two groups of UDS transducers were tested with the cavitometer and with one cleaning test; 48 units of the Jet1 UDS model and 12 units of the Jet2 model (Gnatus, Ribeirao Preto, Brazil) were tested. Sixteen assessments were carried out with each transducer, for a total of 768 measurements with Jet1 and 192 with Jet2. Spearman test was employed to observe correlations between the measurements from the different methods. Jet1 transducers were compared with Jet2 employing Student t test for cavitometer assessments and Mann Whitney U test for the cleaning test assessments. Results: Was found correlation between the cavitometer and cleaning results. The developed cavitometer detected a significant difference between average results for the two transducer groups (Jet1: 4.3±0.7 Vs Jet2: 6.2±0.5) and this was also observed for the cleaning test (Jet1: 7.0±0.8 Vs Jet2: 8.7±0.4) Conclusion: The results showed that the developed cavitometer is adequate for the UDS quantitative performance assessment, particularly for replacement of the cleaning test for industrial quality control.
Highlights
The removal of dental plaque and calculus by ultrasonic devices was first evaluated in the 1950s (Johnson and Wilson, 1957)
The effects of removing dental plaque and calculus with ultrasonic devices have been extensively studied (Zitterbart, 1987; Guentsch and Preshaw, 2008; Walmsley et al, 2008) and in this context, the removal efficiency is attributed to the chipping action of the probe tip and cavitation activity induced in the ultrasonic transducer cooling water (Walmsley et al, 1984; Drisko, 1998; Oda et al, 2000; Arabaci et al, 2007)
The focus has primarily been in two directions: Measuring the tip vibration amplitude and measuring the cavitation intensity induced by the tip
Summary
The removal of dental plaque and calculus by ultrasonic devices was first evaluated in the 1950s (Johnson and Wilson, 1957). It is a widely accepted clinical procedure and probably the most commonly recognized application of power ultrasonic in dentistry, followed by ultrasonic cleaning baths. The effects of removing dental plaque and calculus with ultrasonic devices have been extensively studied (Zitterbart, 1987; Guentsch and Preshaw, 2008; Walmsley et al, 2008) and in this context, the removal efficiency is attributed to the chipping action of the probe tip and cavitation activity induced in the ultrasonic transducer cooling water (Walmsley et al, 1984; Drisko, 1998; Oda et al, 2000; Arabaci et al, 2007).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.