Abstract

The processes limiting the accuracy or range of total pressure measurements in the UHV range by ionization gauges are examined. These processes include x-ray induced photoemission, ion currents resulting from electron stimulated desorption, Kingdon orbits of the ions in Bayard–Alpert gauges, variations in electron yield by ion impact at electrode surfaces, and the effects of variations in electrode geometry. The various methods and designs of UHV ionization gauges that have been developed to reduce or eliminate these limiting processes are critically reviewed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.