Abstract
In this study, a cascade screening system has been developed combining Dual Super Polymerase Chain Reaction (DSPCR) with the universal Lateral Flow Biosensor (LFB) for the ultrafast, universal and visual screening of dual GM elements, taking P-35s × T-nos for example. In the design of DSPCR for universal screening, gene-specific forward primers were labelled with biotin and gene-specific reverse primers were tagged with Cy5 and digoxin, respectively. In 2.5-min, DSPCR effectively amplified the dual target fragments through our prototype facility. Then, through specific antigen-antibody binding, a universal lateral flow biosensor exported visually dual-amplified results simultaneously without cross contamination. After optimization, the detection limit allowed 0.05% GM maize, corresponding to nine copies in maize. The entire detection process could be achieved in 10 min without any large-scale instrumentation. This method may be useful for the ultrafast, universal and visual screening of dual GM elements (P-35s × T-nos) in GM crop lines and is expected to be of great promise for rapid GMO screening and point-of-care tests.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.