Abstract

This paper introduces a new version of a fast and accurate O(N3logN) reconstruction algorithm for cone-beam CT with a circular source trajectory. This is the first report on the performance of this algorithm with high-resolution large-matrix Micro-CT data. We determined that a software implementation of the FHBP algorithm on off-the-shelf workstations can provide speedups by an order of magnitude or more in Micro-CT reconstruction with no perceptible loss of image quality as compared to state-of-the-art commercial software implementations of conventional reconstruction. This is especially important for Micro-CT reconstructions, where maximum image size can be as large as 8 Ktimes8 K pixels per slice and reconstruction of such image volumes can span days. Owing to its algorithmic acceleration, the FHBP-based code achieves the performance of a cluster of computers on a single PC, and could provide even faster reconstruction rates if running on the entire cluster.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.