Abstract
Separation speed and resolution are two important figures of merit in chromatography. Often, one gains the speed at the cost of the resolution, and vice versa. Scientists have employed short-packed columns for ultrafast separations but encountered challenges such as limited mobile phase velocity, extra-column effect caused band broadening, and column packing difficulty. We have recently demonstrated ultrahigh resolutions of narrow open tubular liquid chromatography (NOTLC); this allows us to trade some of the resolution for speed. In this work, we explored NOTLC for ultrafast LC separations. We used a 2.7 cm (effective length) narrow open tubular (NOT) column and showed a baseline separation of 6 amino acids in less than 700 ms. Ways to further increase the speed were discussed. Using short narrow open tubular (NOT) columns to perform ultrafast separation we overcame the challenges from using short packed columns. To demonstrate the feasibility of using this ultrafast separation technique for practical applications, we separated complex protein digests; peptides were nicely resolved in ∼1 min.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.