Abstract

The XAFS station on beamline U7C of National Synchrotron Radiation Laboratory (NSRL) was completely constructed in December 1998. The source for XAFS station is from a 3-pole superconducting wiggler with magnetic field of 6 T inserted in the straight section of the storage ring. Using a Si(111) double crystal monochromator with a fixed slit, the X-ray intensity at the sample position is about 3x10(9) photons/second at the energy of 8980 eV of Cu K-edge. The Keithley 65 17 electrometers are used to record the electron charges that are produced in the ionization chambers. A high ratio of signal to noise has been obtained for the XAFS spectra of Cu, Ni and Fe foils. Furthermore, the XAFS spectrum of Cu foil in NSRL is in good agreement with that obtained in BSRF and KEK.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.