Abstract

The capability of retrieving spectral information from line shapes recorded by two-tone frequency-modulation spectroscopy (TTFMS) is investigated. A TTFMS theory accounting for dispersion and nonlinear distortion of diode laser frequency modulation response is presented. The adequacy of the theory for a detailed modeling of line shapes recorded with high resolution is examined. An extensive error analysis of line parameters (i.e., width, intensity, and line center) retrieved by a nonlinear least-squares fitting procedure is made. Plots of residual errors with characteristic signatures that are due to incorrectly assigned modulation parameters and choice of line profile are presented. In least-squares fits to experimental oxygen data with a Voigt profile influence from collisional(Dicke) narrowing is clearly exhibited, and when we used a collisionally narrowed line profile deviations of the model were reduced to less than 0.2%. We demonstrate that accurate quantitative measurements by TTFMS over a wide range of concentrations, temperatures, and pressures are possible.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.