Abstract

A high resolution tunable infrared diode laser spectroscopy system was developed for isotope analysis with sensitivity at ppb levels. Such a system is ideally suited for detection and measurement of minute amounts of infrared active compounds present in a huge noninfrared active background such as air. The operation and capabilities of the system were demonstrated by measuring physiological levels of isotopic carbon monoxide, 12C16O and 13C16O, naturally present in exhaled human breath with essentially no sample preparation. The simplicity in obtaining such data suggests that fundamental physiological information may be derived from noninvasive measurements. This makes the system potentially useful for many biomedical applications.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.