Abstract

Frequency modulation spectroscopy through a single-mode optical fiber has been demonstrated for monitoring H2O vapor at 7665 cm−1. Using a 2.5 mW distributed feedback laser and 100 m of optical fiber, a minimum detectable absorption of 5×10−7 was achieved. To obtain this sensitivity limit, very large spurious signals and technical noise were suppressed using a dual-channel detection scheme.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.