Abstract

Tri-component, mobile bearing, uncemented, total ankle replacements were introduced after the high failure rates of cemented, highly constrained, first-generation, total ankle replacement implants. A total of 30 primary total ankle replacements in 29 patients (20 males and 9 females) were followed up in the present retrospective study for up to 13 (mean 5.1 ± 4) years. The postoperative functional and radiographic outcomes were measured. Failure was defined as revision of either of the components for any reason or conversion of the total ankle replacement to arthrodesis because of debilitating pain that did not resolve after surgery. Of the 29 patients, 2 underwent revision and 1 underwent arthrodesis. All 3 patients had the malpositioned talar implant revised. The mean American Orthopaedic Foot and Ankle Society score was 81 at 1 year postoperatively. Revision of the tibial or talar component for any reason or conversion of the ankle replacement to arthrodesis was considered failure for the survival analysis. Kaplan-Meier analysis showed a 5-year survival rate of 87.6%. The last failure occurred 23.3 months after surgery.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.