Abstract

The purpose of this study was to compare the rate of proximal type I endoleaks in patients undergoing endoluminal repair of infrarenal abdominal aortic aneurysms with endografts having either transrenal or infrarenal fixation. From September 1998 to May 2000, 42 patients received endoluminal aortic grafts for the treatment of infrarenal abdominal aortic aneurysms. Patients received either transrenal or infrarenal devices, based on the surgeon's preference. All patients had infrarenal aortic neck lengths measuring at least 1.5 cm and proximal neck angulation of <60 degrees. The endoluminal grafts were oversized by 10-20% relative to the diameter of the infrarenal aorta. The presence of endoleaks was determined at the initial procedure by contrast angiography and during subsequent follow-up at 1 month and 3 months by CT scan or duplex ultrasound. No significant differences in the rate of proximal type I endoleaks can be demonstrated between transrenal and infrarenal device types in this small cohort. Proper patient selection is more important than type of proximal fixation in preventing endoleaks.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.