Abstract

The relationship between peak airway pressure, alveolar pressure and respiratory frequency was calculated for the range of compliances and airway resistances which might be encountered during mechanical ventilation of a 3-kg neonate. The pressure/flow relationships of 2.5, 3.0, 3.5 and 4-mm tracheal tubes were determined at a series of flows from 0.5 to 4 litres/minute. Peak airway and alveolar pressures were then measured at various frequencies and inspiratory:expiratory ratios with the tubes incorporated in a model lung. Large differences between peak airway and alveolar pressures developed when frequency was increased or inspiratory time decreased; the differences were greatest with the smaller tubes. Shortening expiratory time by increasing the frequency or altering the inspiratory:expiratory ratio resulted in increased end-expiratory pressure because of incomplete emptying of the lung.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.