Abstract

Recently, guided surgery has been introduced and several papers verified its accuracy. While those studies reported on the accuracy of the entire procedure, this experiment wanted to make clinicians aware of the amount of the deviation that can already occur during the drilling procedure alone, due to the tolerance of the drill in the drill key. Drilling was executed in plexi-glass with a maximal inclination of the drills within the drill keys. A mean deviation in angulation of 4.7 degrees occurred with a mean horizontal deviation at the implant shoulder of 0.8 mm and 1.8 mm at the apex of 13 mm implant. The deviation was further dependent of the implant length, the distance of the sleeve above the bone and the respective guiding system. Increasing the height of the drill key will minimize the inaccuracy. The results from this experiment showed the importance of keeping the drill parallel to the guide in a centric position.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.