Abstract
Gamma-ray energy resolution of a NaI(Tl) inorganic scintillation detector was improved in the present work. For this purpose, leading edge and constant fraction timing methods were used. Gamma energy spectrum of the radioisotope (<SUP>137</SUP>Cs) and the time spectra through these timing methods were obtained separately. Energy spectrum was gated with both time spectra, and the obtained results were compared with each other. Energy resolution enhancements of 3.5% and 7.5% on the photopeak of the radioisotope were obtained through leading edge and constant fraction timing methods, respectively. Energy resolution value from constant fraction timing was better than that of leading edge timing method.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: International Journal of Instrumentation Science
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.