Abstract

A new diffusion-based solid-phase microextraction (SPME) time-weighted average (TWA) field water sampling device was developed and investigated by field trial. The sampler is constructed with copper tube and caps and a commercial SPME fiber assembly. The device possesses all advantages of SPME; it is solvent-free, reusable, combines sampling, isolation and enrichment into one step, and the fiber can be directly injected into a gas chromatograph for analysis with a commercial SPME fiber holder, without further treatment. Field trials in Laurel Creek (Waterloo, Ont., Canada) and Hamilton Harbour (Hamilton, Ont., Canada) illustrated that the device is durable, easy to deploy, and the mass uptake of the device is independent of the face velocity. The device provides good precision [relative standard deviations (RSDs) are less than 20%] and the data obtained with this device are quite comparable to those obtained with the spot sampling method, which demonstrates that the newly developed SPME water sampling device is suitable for long-term monitoring of organic pollutants in water.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.