Abstract

We investigated the correlation between hepatic venous pressure gradient (HVPG) and liver shear wave speed (SWS) measured by multi-frequency time-harmonic ultrasound elastography (THE) before and after transjugular intrahepatic portosystemic shunt (TIPS) implantation. Ten patients with ascites, cirrhotic liver disease and portal hypertension were prospectively examined with invasive HVPG measurement and THE before and after TIPS implantation. HVPG and SWS decreased after TIPS placement from 20.4 ± 2.2 mmHg to 9.8 ± 4.1 mmHg (mean ± standard deviation) and from 3.87 ± 0.54 m/s to 3.27 ± 0.44 m/s. Mean reduction HVPG was −10.6 ± 3.7 mmHg, p < 0.001; mean reduction SWS was −0.60 ± 0.29 m/s, p < 0.001. A linear correlation was observed between HVPG and SWS (R = 0.59, p = 0.0061). THE-measured SWS is a first potential direct ultrasound marker for liver decompression following TIPS in ascites-associated cirrhotic liver disease and therefore might be suitable to non-invasively detect portal hypertension.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.