Abstract

Purpose: To describe a technique to catheterize antegrade branches of a branched thoracoabdominal endograft from a femoral access with the help of standard sheaths and a vascular suture. Technique: The technique is demonstrated in a patient who underwent successful complex thoracoabdominal branched endovascular aortic repair. After the deployment of an aortic endograft with two antegrade branches for the targeted renovisceral vessels, a standard braided sheath was preloaded with a 3/0 polypropylene suture and introduced inside an additional sheath from the groin to the thoracic aorta. Simultaneous gentle traction on the suture as the preloaded sheath was advanced achieved a very stable 180° curve of the proximal end of the sheath. It was possible to selectively catheterize the antegrade branches and respective target vessels sequentially, as well as deploy the planned bridging stents for each branch. Conclusion: The through-and-through suture technique is a helpful tool in branched endovascular aortic repair. It saves time, radiation, and materials; no snare is needed, and it can be preloaded into a sheath.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.