Abstract
Reproducibility of the determination of Young's modulus and energy absorption along the three axes of trabecular bone cubes was analysed by non-destructive compression to 0.5% strain using different testing protocols. These protocols included testing with and without pre-conditioning to a viscoelastic steady state, and different orders of test directions. Reproducibility of conditioned tests was generally better than that of non-conditioned tests. No major effect of changing the order of the test direction was found. Three-axial conditioned testing of cubes from the proximal tibial epiphysis of five humans revealed a global transverse isotrophy while most cubes showed orthotropy. The ratio between stiffness along the long axis of the tibia and the stiffness in the transverse plane was 3.7 ± 0.4 (mean ± 2 SE). The corresponding ratios for elastic energy storage and viscoelastic energy dissipation were 2.5 ± 0.2. There was no difference between the relative energy loss during a testing cycle (loss tangent) in the three axes.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.