Abstract

Thin-film platinum resistance thermometers (PRTs) are generally manufactured using the deposition of a thin platinum film on an alumina substrate and a laser-trimming method. Because of the strong adhesion between the platinum thin film and the alumina substrate, the PRTs inevitably have strain over the operating temperature range. This causes anomalies and instabilities in the resistance versus temperature characteristics (R–T). The most prominent and observable effect of thermally induced strain is the thermal hysteresis in the R–T characteristics. Thermal hysteresis is one of the main uncertainty factors in the calibration of industrial platinum resistance thermometers in laboratories. The thermal hysteresis for 30 thin-film PRTs was measured in the range of 0 °C to 500 °C in 100 °C steps. The thermal hysteresis was measured repeatedly using the same process, and the hysteresis decreased drastically with the repeated measurements. The thermal hysteresis was distributed from 16 mK to 156 mK for all sensors, and the lowest hysteresis was 1 mK to 11 mK in the test temperature range.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.