Abstract

The LCG Limited organic DSC calibration standards have been used to carry out the enthalpy calibration of a power compensated DSC in a temperature range (42–147°C) for which suitable metal standards are not available. The mean calibration factor was in agreement with the value obtained using indium, thus demonstrating that there was no significant error in using indium as a single point calibrant for measurements on the fusion of organic materials at lower temperatures.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.