Abstract

This paper describes a simple method for the determination of sixteen elements in food samples by using inductively coupled plasma spectrometry (ICP-MS). Prior to analysis, 100–250 mg of powdered food samples were accurately weighed into a Teflon digestion vessel. Then, 4 ml of 20% v/v concentrated nitric acid and 2 ml of hydrogen peroxide were added. Decomposition of samples was carried out in a microwave digestion system. In order to verify the accuracy and precision of the proposed method, five Standard Reference Materials from the National Institute of Standards and Technology (NIST) (Whole Egg Powder RM 8415, Rice Flour SRM 1568a, Typical Diet SRM 1548a, Wheat Flour SRM 1567a and Bovine Muscle Powder RM 8414) were analyzed. Additional validation data are provided based on the analysis of 18 different types of food samples by the proposed method and using comparative methods with AAS as the detector.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.