Abstract
During the administration of therapeutic ultrasound, the amount of pressure at the sound head-tissue interface may affect the physiological response to and the outcome of treatment. Speed of sonification; size of the treatment area; frequency, intensity, and type of wave; and coupling media are important parameters in providing the patient with an appropriate ultrasound treatment. Pressure variations affect ultrasound transmissivity, yet pressure differences have been virtually unexplored. The purpose of this study was to assess the effects of sound head pressure on acoustic transmissivity. Three trials were conducted whereby pig tissue was subjected to increased sound head pressures using manufactured weights. The weights were added in 100 g increments, starting with 200 g and finishing with 1,400 g. Increased pressure on the transmitting transducer did affect acoustic transmissivity; acoustic energy transmission was increased from 200 g (0.44 lb) up to and optimally at 600 g (1.32 lb). However, there was decreased transmissivity from 700 to 1, 400 g (1.54 to 3.00 lb).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.