Abstract

The problems that arise during the calibration of the reference plates of large-aperture interferometers are discussed and the methods for solving them are presented. It is shown that the repeatability of shapes of the reference plates after their rearrangements can be provided by selecting the mechanism of their fixing in place. It is demonstrated that the temperature deformations of the reference-plate surfaces can be determined from a numerical model. These measures have made it possible to calibrate the reference plates of the 630-mm interferometer with an accuracy of λ/1000 RMS (λ = 632.8 nm).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.