Abstract

Objective: To determine the effectiveness of objective statistical detection in CAEP testing to evaluate audibility in young infants with sensorineural hearing loss. Design: CAEP recordings to speech-based stimuli were made at three presentation levels (55, 65, or 75 dB SPL) when a group of hearing-impaired infants were either aided or unaided. Later-obtained behavioral audiograms were used as the gold standard against which to evaluate the accuracy of the automatic detection of the presence/absence of CAEP responses. Study sample: Participants were 18 infants with confirmed sensorineural hearing loss. Results: Higher sensation levels led to a greater number of present CAEP responses being detected. More CAEP waveforms were detected in the aided condition than in the unaided condition. Conclusion: Our results suggest that the presence/absence of CAEP responses defined by the automatic statistical criterion was effective in showing whether increased sensation levels provided by amplification were sufficient to reach the cortex. This was clearly apparent from the significant increase in cortical detections when comparing unaided with aided testing.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.