Abstract
ABSTRACTPurpose: To compare the performance of the PlusoptiX S12 mobile photoscreener and the Retinomax K+3 Autorefractor as screening devices in preschool children.Methods: Children ranging from 3 to 5 years of age from 11 San Diego County preschools underwent vision screening in their schools where ambient light could not always be controlled using both the Retinomax and the PlusoptiX. Cycloplegic refraction on the consented children was subsequently performed on the UCSD EyeMobile for children on-site at the school locations.Results: A total of 321 children were screened with the PlusoptiX and Retinomax. The PlusoptiX referred 22% of children, of whom 70% of the referrals were read as “unable”. The Retinomax referred 13% and there were no “unables”. Similar results occurred in the cycloplegic-refracted 182 consented children—64% of the PlusoptiX referrals were read as “unable” . Only one third of these “unables” required glasses. Both devices referred the four children with amblyopia and one case of st...
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.