Abstract

This study determines the flammability of polyvinylchloride (PVC), red rubber (RR), and silicone (Si) endotracheal tubes in oxygen- and nitrous-oxide-enriched atmospheres. Flammability is measured by using the oxygen and nitrous oxide indices of flammability with laser ignition. The laser-ignited oxygen (O2) index of flammability of the endotracheal tubes is: PVC, 0.25; RR, 0.19; Si, 0.20. The laser-ignited nitrous oxide (N2O) index of flammability of the endotracheal tubes is: PVC, 0.45; RR, 0.37; and Si, 0.41. These results are similar to the previously reported O2 and N2O indices of flammability with propane-torch ignition. This study validates the concept that the indices of flammability are useful measures of endotracheal tube flammability and are independent of the ignition source.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.