Abstract
A novel method was introduced for the quantitative determination of substances in aqueous solutions by using the evaporative light scattering (ELS) detector of a high performance liquid chromatograph (HPLC). The principle of the measurement is the different equilibrium vapor pressure of the solvent and the analyte resulting in decreasing evaporation rate, larger droplets and stronger signal with increasing concentration. The new technique based on vapor pressure analysis was validated with traditional UV-Vis detection carried out with a diode array detector (DAD). The new technique was used for monitoring the concentration of solutions obtained during the enzymatic degradation of poly(3-hydroxybutyrate) yielding the 3-hydroxybutyrate monomer as the product. The accuracy of the measurement allowed the determination of degradation kinetics as well. The results obtained with the two techniques showed excellent agreement at small concentrations. Deviations at larger concentrations were explained with the non-linear correlation between analyte concentration and detector signal and the linear regression used for calibration. Mathematical analysis of the method made possible the determination of the evaporation enthalpy of the analyte as well. The new approach is especially suitable for the quantitative analysis of compounds, which do not absorb in the detection range of the DAD detector or if their characteristic absorbance is close to the lower end of its wavelength range.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.