Abstract
Limiting the risk of dislocation is one of the main aims of both dual mobility and Lefèvre retentive cups. However, these devices have never been compared. The goal of our study was to compare these devices in total hip arthroplasty revisions for instability. The judgement criterion was non-recurrence of dislocation in a follow-up period of eight years. This retrospective case-control study compared two continuous paired series of total hip arthroplasty revisions for instability. These series included 63 patients and 159 patients with implantation of a Lefèvre retentive cup and a dual mobility cup, respectively. The success rate at eight years (i.e., no recurrence) was 91 ± 0.05% and 95 ± 0.02% in the Lefèvre retentive cup and dual mobility groups, respectively. The difference was not statistically significant (p = 0.6). It seems that the Lefèvre retentive cup provides comparable outcomes with the dual mobility cup in the total hip arthroplasty revisions for instability, avoiding recurrence in long term.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.