Abstract

The Time Projection Chamber (TPC) is the core tracking detector for the STAR experiment at RHIC. To determine spatial distortions, calibrate and monitor the TPC, a laser calibration system has been built. We developed a novel design to produce ∼500 thin laser beams simulating straight particle tracks in the TPC volume. The new approach is significantly simpler than the traditional ones, and provides a higher TPC coverage at a reduced cost. During RHIC 2000 and 2001 runs the laser system was used to monitor the TPC performance and measure drift velocity with ∼0.02% accuracy. Additional runs were recorded with and without magnetic field to check E×B corrections.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.