Abstract
The jackknife as an approach for uncertainty estimation in gamma spectrometric uranium isotope ratio measurements was evaluated. Five different materials ranging from depleted uranium (DU) to high enriched uranium (HEU) were measured using gamma spectrometry. High resolution inductively coupled plasma mass spectrometry (ICP-SFMS) was used as a reference method for comparing the results obtained with the gamma spectrometric method. The relative combined uncertainty in the gamma spectrometric measurements of the 238 U/ 235 U isotope ratio using the jackknife was about 10–20% ( k = 2), which proved to be fit-for-purpose in order to distinguish between different uranium categories. Moreover, the enrichment of 235 U in HEU could be measured with an uncertainty of 1–2%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.