Abstract
Experimental and finite element investigations were carried out on axisymmetric models of the femoral component of a total hip replacement. In one instance, the interface between the stem and the surrounding bone cement was assumed to be rigidly bonded; in a second, it was allowed to slip. For the latter case, a friction coefficient of 0.2 was determined experimentally. The predictions of the finite element models demonstrated excellent agreement with the results from the experimental tests at all sites where comparisons were made, thus validating these models. The effect of stem-cement slip was shown to reduce the maximum shear stress in the cement mantle by approximately 30 per cent.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.