Atrial spatial vectorcardiograms (P sÊ-loops) were recorded for 20 normal individuals using high gain amplifiers with different halfpower upper frequency limits. The effects of deep inspiration and expiration were also recorded, showing that maximal respiratory movements alter the P sÊ-loop in an unpredictable fashion. Differences in the maximal frequency response of the vectorcardiograph influenced the configuration of the P sÊ-loop recorded. It is suggested that a maximal frequency response of 50 Hz is satisfactory for the recording of P sÊ-loops for clinical and some research purposes.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.