Abstract

Computational fluid dynamics has been widely used to assist the design and evaluation of blood pumps. Discretization errors associated with computational grid may influence the credibility of numerical simulations. Non-conformal grid interfaces commonly exist in rotary machines, including rotary blood pumps. Should grid size across the interface differ greatly, large errors may occur. This study explored the effects of non-conformal grid interface on the prediction of the flow field and hemolysis in blood pumps using large eddy simulation (LES). Two benchmarks, a nozzle model and a centrifugal blood pump were chosen as test cases. This study found that non-conformal grid interfaces with considerable change of grid sizes led to discontinuities of flow variables and brought errors to metrics such as pressure head (7%) and hemolysis (up to 14%). The results on the full unstructured grid are more accurate with negligible changes of flow variables across the non-conformal grid interface. A full unstructured grid should be employed for centrifugal blood pumps to minimize the influence of non-conformal grid interfaces for LES simulations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.