Abstract

These investigations were performed to determine histologic and incisional consequences of varying pulse duration, duty cycle, and average powers during laser incision at 9.3 μm in soft tissue. In 19 fresh pigs' jaws six standardized incisions 3 cm long were made per parameter with a template and motorized jig. Laser parameters investigated were average power: 1 to 9 W, duty cycle: 10% to 80%, and pulse duration: 1 to 200 msec. The gated Cw mode was used. Incision width and depth and collateral tissue effects were assessed statistically with general linear procedures. Multiple factors were found to influence the outcome of laser irradiation. Depth of incision correlated positively with average power. Tissue damage correlated strongly and negatively with all three variables. These results demonstrate that a wide range of surgical and collateral effects can be achieved with one specific laser device depending on the parameter configuration selected.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.