Abstract

Background: Intracoronary devices such as stents or guide wires may disturb the dose distribution of beta sources in cardiovascular brachytherapy. As clinical observations indicate that underdosage increases the risk of restenosis, accurate measurements are mandatory to investigate these effects. Methods and results: Dose perturbation effects of different interventional equipment were systematically determined. Dose distributions of 90Sr-beta line sources were measured by means of a special set-up employing plastic scintillator dosimeters in a water phantom. Shielding effects were found to be 2–5% for single stents and 5–10% for graft stents, stent–in-stent geometries, and guiding catheters. Guide wires close to the source reduced the dose by 25–30%. Conclusions: Beta dose perturbation effects of typical stent types are almost negligible and can be corrected by an increased source dwell time if necessary. Guide wires produce effects which are clinically much more important and should therefore be retracted from the irradiation area.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.