Abstract

AbstractA new device setup for temperature and injection‐dependent lifetime spectroscopy (TIDLS) is described. It comprises two off‐the‐shelf components: a heating and cooling stage (HCS) from INSTEC and an inductive‐coil photoconductance (PC) instrument (WCT‐100) from Sinton Consulting Inc. The HCS was fitted to the WCT‐100 in a manner that circumscribes the inductive coil (the sensor) of the RF bridge circuit and controls the temperature of the wafer effectively. This setup has the advantage of requiring minor modifications to industry standard instruments while attaining a large temperature range. As experimental verification, injection‐dependent lifetimes were measured over a temperature range, 0–230°C, in three iron‐implanted silicon wafers. The measured lifetimes are consistent with the Shockley–Read–Hall equation using the impurity concentration calculated from the implant dose and the energy level and capture cross‐sections of interstitial iron from the literature. Copyright © 2008 John Wiley & Sons, Ltd.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.