Abstract

Abstract Background and purpose This study investigates the influence of an MRI linac fringe field on neighbouring radiotherapy accelerators and whether restoring their original steering calibration can be used to shorten downtime after changes of the magnetic field of the MRI linac. Methods and materials A 1.5 T MRI linac was installed in a treatment room, neighboured by three accelerators. A detector array was used to acquire gantry angle dependent beam flatness and symmetry parameters. Measurements were performed before and after ramping up the magnet, and again after adjusting the beam steering for deviations in flatness and symmetry. Results Changes in beam symmetry as a function of gantry angle of up to 4% (Linac A) and 1% (Linac B) were observed, causing beam parameters to be out of clinical tolerance. After steering adjustment, the parameters were within tolerance for all gantry angles. Linac C did not show any significant change. Conclusion The MRI linac magnetic fringe field can influence the beam steering of the accelerators in adjacent bunkers. The steering of two accelerators that were situated the closest to the MRI linac needed to be adjusted in order to bring beam parameters within tolerances. Adjusting the steering completely compensated the influence of the magnetic fringe field of the MRI linac.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.