Abstract
An instrument called the paste viscosity indicator (PVI) has been developed at the Grain Research laboratory (GRL) for measuring the paste viscosity of a heated ground grain-water slurry. Increased resistance offered to the stirring device upon heating a ground grain-water slurry is reflected in a voltage signal representing the current requirements of a small DC motor. Paste viscosity decreases with increasing α-amylase levels in the ground grain. The viscosity is measured by a digital panel meter which retains its reading after a testing period of 3 or 3.5 min. The machine is portable (8.8 kg) and does not require a running water source. Operation is in simple steps and sample containers are low cost and disposable. The method correlates well with Falling Number method over the range of 65 – 425 s for Canadian Western Red Spring wheat. As such, it may be well-suited for implementation into a carlot or cargo monitoring program for assessment of preharvest sprout-damage.Key words: Amylase, wheat, control, sprout-damage
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.