We report here on our latest intercomparison of the Josephson junction noise thermometer and 60Co γ-ray anisotropy thermometer from 0.01 K to 0.05 K. These results show both thermometers are in agreement to better than 0.5%. This is an improvement over our previous data and is due mainly to better temperature control and advances in instrumentation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.