Abstract
The esophageal obturator airway (EOA) has been approved for use in cardiopulmonary resuscitation (CPR) since 1973. Complications reported with its use include esophageal rupture, aspiration, and inadvertent tracheal occlusion or intubation with the obturator. Five cases of unrecognized tracheal intubation occurring in 170 cases of EOA use during CPR are reported. Records were analyzed for extent and treatment of injuries in the field and the hospital, blood gas levels before and after replacement of the erroneously placed EOA with an endotracheal tube (ETT), and outcome. Blood gas analysis confirmed absent ventilation and profound acidosis following tracheal intubation with the EOA with average EOA arterial blood gases: pH 6.9/pCO2 78/pO2 47. Mortality from this complication was 100%. Tracheal intubation with the EOA is under-recognized and under-reported. It is a potentially lethal complication. The EOA should be replaced with standard bag mask technique or preferably endotracheal intubation in the field by emergency medical technicians.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.