Abstract

Bone mineral mass measurements using dual photon absorptiometry rely on the assumption that fat thickness is constant within the region scanned by the photon beam. This assumption has been tested using both single and multiple slices from archived CT scans. In 26 patients, the difference in fat content between the bone and baseline measurement sites, the fat deviation, was found to vary from -2.7 to 18.7%. The median fat deviation was 4.4% which can lead to errors in measured lumbar spine bone mineral of 6 to 11%. Approximately one third of patients will have errors greater than 10%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.