Abstract
The purpose of this study was to determine the effect of patellar bone and patellar implant thickness on patellofemoral forces after resurfacing in total knee arthroplasty. Seven cadaver knees were tested using an Oxford Knee Testing Rig. This model gave the specimens 6 degrees freedom while dynamic data were collected. Each knee was tested from full extension to 95 degrees knee flexion. Knees were implanted with Press Fit Condylar femoral, tibial, and patellar implants. The effect of varying patellar thickness on patellofemoral forces was determined by using custom modular oval-domed polyethylene patellar components with progressive thickness increments of 2 mm. Patellofemoral forces were measured by a custom-designed uniaxial patellar load cell. Statistically significant increases in patellofemoral compression forces were found from 70 degrees to 95 degrees flexion with increased patellar bone and patellar implant thickness.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.