Abstract

A simplified finite element model of the human lumbar intervertebral disc was utilized for understanding nucleus pulposus implant mechanics. The model was used to assess the effect of nucleus implant parameter variations on the resulting compressive biomechanics of the lumbar anterior column unit. The effects of nucleus implant material (modulus and Poisson's ratio) and geometrical (height and diameter) parameters on the mechanical behavior of the disc were investigated. The model predicted that variations in implant modulus contribute less to the compressive disc mechanics compared to the implant geometrical parameters, for the ranges examined. It was concluded that some threshold exists for the nucleus implant modulus, below which little variations in load-displacement behavior were shown. Compressive biomechanics were highly affected by implant volume (under-filling the nucleus cavity, line-to-line fit, or over-filling the nucleus cavity) with a greater restoration of compressive mechanics observed with the over-filled implant design. This work indicated the effect of nucleus implant parameter variations on the compressive mechanics of the human lumbar intervertebral disc and importance of the "fit and fill" effect of the nuclear cavity in the restoration of the human intervertebral disc mechanics in compression. These findings may have clinical significance for nucleus implant design.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.