Abstract

scatter photon produced during radiationtherapy with high energy photons is the main source of unwanted out-of-field and superficial received doses of patients.Surface buildup dose is dependent on electron contamination primarily from the unblocked view of the flattening filter and secondarily from air and collimation systems .We performed a comprehensive set of surface and buildup dose measurements with a thin window parallel-plate (PPC-40) chamber to examine effects of attenuating media in front of 6 MV X ray. To evaluate the impact of beam segmentation on buildup dose, measurements were performed with 10 × 10 cm2 fields, Measurements were performed in Solid Water using parallel plate chambers and diode for a 6 MV X-ray beam.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.