Abstract

The spatial distribution of induced currents from the stimulating coil is calculated from a computer model with graphical output. Two configurations of a plane circular coil are considered: parallel to the tissue surface and perpendicular to the surface. The surface is assumed planar and infinite in extent. The tissue is modelled as a uniform, isotropic volume conductor. A quasi-static approximation is made in calculating the electric field. Maps of current density, J, as a function of position, including depth, are shown. In both configurations, J is always parallel to the surface, and is maximum at the surface. There is no perpendicular (vertical) current. For a one-turn 10 cm diameter coil, spaced 1 cm from conducting tissue and parallel to it, with rate of change of current 108 A s-1, Jmax=6.8 A m-2 (assuming conductivity 0.2 Omega -1 m-1). In the perpendicular configuration Jmax=4.1 A m-2. These results suggest that nerve fibres running parallel to the skin surface are more likely to be stimulated than those running obliquely; and that it is extremely difficult to stimulate nerve fibres running perpendicularly.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.