Abstract
Abstract (1)The effect of light on the fluorescence of the ethylene diamine condensation products of the oxidized catechol amines has been studied. (2)It was found that exposure of the norepinephrine derivative for 30 minutes to light reduced the fluorescence to a very low level. The epinephrine complex, on the other hand, is only slightly affected by the same exposure. (3)On the basis of this differential sensitivity to light a sensitive method for the quantitative determination of the two catechol amines in blood has been developed. (4)The levels of epinephrine and norepinephrine in the blood of normal individuals were found to range from 0.1-1.5 µg per liter and 0.3-4.0 µg per liter, respectively. (5)Further work must be done before this method can be applied to the determination of the catechol amines in urine.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.