Abstract
Twenty-six patients with refractory ventricular arrhythmias received the automatic implantable cardioverter-defibrillator. A patch lead only was placed during arrhythmia surgery in 7 other patients. During 13 +/- 6 (SD) months, the device discharged in 10 patients because of a sustained ventricular arrhythmia. No sudden deaths occurred. There were 31 complications in 17 patients, including postoperative refractory heart failure, coronary artery erosion, subclavian vein thrombosis, postoperative stroke after conversion of atrial fibrillation, atelectasis with pneumonia, symptomatic pleural effusions, and infection at the generator site. The cardioverter-defibrillator discharged in 9 asymptomatic patients, failed to terminate ventricular fibrillation during postoperative testing in 3 patients, and had premature battery failure in 4 patients. Tachycardia slowing during chronic amiodarone therapy and unipolar ventricular pacing during ventricular fibrillation precluded or delayed arrhythmia sensing. Thus, the cardioverter-defibrillator can be life saving, but its potential complications and interactions with antiarrhythmic drugs and pacemakers must be considered at patient selection.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.